LIFE Corporation
 

Refill Information

The LIFE® emergency oxygen units must be filled with medical grade oxygen (USP) if the indicator on the gauge is less than full, in order to ensure that an adequate supply is on hand for emergency.

THE OXYGEN CYLINDER MAY BE REFILLED WITH PURE MEDICAL OXYGEN BY ANY MEDICAL/INDUSTRIAL GAS DISTRIBUTOR who uses the Compressed Gas Association (CGA) #870 PIN-INDEXED UNIVERSAL COUPLING FOR OXYGEN. Gas distributors in your area may be some of the following companies:

  • Airgas
  • Air Liquide Company
  • Air Products Company
  • BOC Gases
  • Linde Gas
  • LifeGas
  • Praxair

(Check your yellow pages under the heading Welding Equipment Gases.)

Both Valves.JPG
CGA870 valves for:
LIFE® SoftPac    LIFE® OxygenPac
LIFE-O2®
LIFE® Start System
 

Pin-Indexed Yoke Connections
Medical gas cylinder valves
for use in respiration therapy,
inhalation or resuscitation applications,
are pin-indexed to prevent
interchangeability with dangerous
or non-compatible gases,
in conformance with CGA standards.

Pin Index Chart BOLD.JPG
Oxygen 2.JPG & 5.JPG
CGA Connection No. 870
USP MEDICAL OXYGEN

NOTE: For LIFE® OxygenPacs ONLY
THE FACTORY SETTING OF THE "ON/OFF" LEVER DOES NOT CHANGE IF THE LEVER REMAINS MATCHED WITH THE RESPECTIVE CYLINDER & VALVE.
The lever must be reset if a unit is reassembled with a different cylinder/valve. The "on" position of each separate cylinder/valve varies anywhere of 360°.
Notice the fingernail activated pin-slide mechanism on the left side of the "on/off" lever. The pin-slide can be retracted so as to set the lever in various positions. The proper position of the lever is so the valve first releases oxygen at the "center" (on the same axis as the contents gauge).
Properly set the lever so that the valve releases oxygen at that position closest to the center of the slot in the Lexan faceplate when assembled.

(It must be remembered that oxygen continues to be released after the valve is turned to "off" for a few seconds, during which the pressure inside the regulator is being released).


email@LIFEcorporation.com

LIFE Corporation, 1776 North Water Street., Milwaukee, WI  53202-1552, USA
Phone 800-700-0202 or 414-272-4000 - Fax 414-272-0000

Products made in the USA.
LIFE® is a Registered Trademark of LIFE Corporation
Site Copyright © 1998- LIFE® Corporation copyright information.

Emergency Oxygen Units Cylinders Regulators Home AHA Emergency Oxygen Guidelines Contact LIFE Refill Information Maintenance Training